Jennings, I
Mackie, I
Arnout, Jef
[UCL]
Preston, FE
We report here results from a United Kingdom National Quality Assessment Scheme (UK NEQAS) exercise in which both plasma spiked with monoclonal antibodies and plasma from a patient known to have lupus anticoagulant (LA) were distributed to 245 hemostasis laboratories with a request for them to test for possible LA using their routine screening procedure. In general, good agreement was seen in the diagnosis of samples spiked with monoclonal antibodies against beta2-glycoprotein 1 (beta2GP1) and prothrombin, the LA-positive patient sample, and a normal pooled plasma; over 87% of centers correctly identified each sample. However, methods employing platelet neutralizing procedures were associated with a higher proportion of false-negative responses with the antiprothrombin-spiked sample, and it is important to recognize that sensitivity and responsiveness of different methods may vary between artificial plasmas and different LA-positive patient plasmas.
- Somers E, J Rheumatol, 29, 2531 (2002)
- Horbach DA, Thromb Haemost, 76, 916 (1996)
- Galli M., Lupus anticoagulants are stronger risk factors for thrombosis than anticardiolipin antibodies in the antiphospholipid syndrome: a systematic review of the literature, 10.1182/blood-2002-02-0441
- Galli M, Thromb Haemost, 84, 1012 (2000)
- Jennings I, Thromb Haemost, 77, 934 (1997)
- Roussi Jacqueline, Roisin Jean Paul, Goguel Alain, Lupus Anticoagulants:First French Interlaboratory Etalonorme Survey, 10.1093/ajcp/105.6.788
- Jennings Ian, Greaves Mike, Mackie Ian J., Kitchen Steve, Woods Timothy A. L., Preston F. Eric, , Lupus anticoagulant testing: improvements in performance in a UK NEQAS proficiency testing exercise after dissemination of national guidelines on laboratory methods, 10.1046/j.1365-2141.2002.03821.x
- Ichikawa K, Lupus, 1, 239 (1992)
- Takeya H, J Clin Invest, 99, 2260 (1997)
- Arnout J, Thromb Haemost, 81, 929 (1999)
- Kitchen S, Thromb Haemost, 72, 426 (1994)
- Exner T, Br J Haematol, 40, 143 (1978)
- Greaves M., Cohen H., Machin S. J., Mackie I., Guidelines On The Investigation and Management Of The Antiphospholipid Syndrome, 10.1046/j.1365-2141.2000.02069.x
- 14 Medical Devices Agency. Evaluation of Lupus Anticoagulant Kits. MDA/98/48 1998. London, UK; Department of Health; 1998 .
- Ames P. R. J., Iannaccone L., Iasio R. De, Brancaccio V., Improved confirmation of weak lupus anticoagulants by employing sensitive and insensitive reagents to the lupus anticoagulant : , 10.1097/00001721-200110000-00009
- Arnout J, Thromb Haemost, 79, 955 (1998)
- Tripodi A, Clin Chem, 49, 1608 (2003)
- Galli M, Blood, 86, 617 (1995)
- Le Querrec A, Thromb Haemost, 86, 584 (2001)
- Arnout J, Br J Haematol, 87, 94 (1994)
Bibliographic reference |
Jennings, I ; Mackie, I ; Arnout, Jef ; Preston, FE. Lupus anticoagulant testing using plasma spiked with monoclonal antibodies: performance in the UKNEQAS proficiency testing programme. In: Journal of Thrombosis and Haemostasis, Vol. 2, no. 12, p. 2178-2184 (2004) |
Permanent URL |