The Carpentier-Edwards bioprosthesis was implanted in 369 patients (414 valves) between May 1977 and December 1987 (age 67.2 +/- 0.5 years); 242 had aortic valve replacement, 80 had mitral valve replacement, 44 had multiple valve replacement, of which 41 were aortic and mitral valve replacement, 2 had isolated tricuspid valve replacement, and 1 had a pulmonary valve replacement. The selection criteria were the following: shorter life expectancy (253 patients) or contraindications to anticoagulants for organic (113 patients) or psychologic (38 patients) reasons, or both. The early mortality rate was 11.1% (aortic valve replacement, 9.1%; mitral valve replacement, 12.4%; aortic and mitral valve replacement, 23.1%). Total cumulative follow-up was 1456 pt-yr (mean 4.4 years, range 1 to 148 months), and the patient evaluation was 99.5% complete. Late mortality was 4.9%/pt-yr. Five-year survival was 70.4% +/- 2.7% overall, 74.3% +/- 3.2% after aortic valve replacement, 60.9% +/- 6.2% after mitral valve replacement (p < 0.03), and 60.7% +/- 8.1% after aortic and mitral valve replacement. Eight patients were reoperated on for primary tissue failure, and freedom from reoperation for structural valve deterioration was 97.5% +/- 1.2% at 5 years and 95.6% +/- 1.8% at 8 years. Failing aortic bioprostheses were explanted in four patients (0.4%/pt-yr) and mitral bioprostheses in seven (1.6%/pt-yr). No patient whose valve was inserted after the age of 70 had to be reoperated on for structural valve dysfunction. The probability of freedom from thromboembolism after 5 and 8 years of follow-up was 93.1% +/- 1.6% and 92.2% +/- 1.8%, respectively. The prevalence of anticoagulant-related hemorrhage was 0.8%/pt-yr (major 0.6%, minor 0.2%). Anticoagulants had to be maintained in 16.3% of the patients: 5.9% after aortic valve replacement, 35.7% after mitral valve replacement, and 45.8% after aortic and mitral valve replacement, while 80.0% were on a regimen of antiplatelet drug therapy. Prosthetic valve endocarditis happened in five patients (0.3%/pt-yr). Freedom from all valve-related morbidity and mortality, including hospital deaths, was 71.0% +/- 2.7% at 5 years and 58.6% +/- 4.6% at 8 years and was significantly better in the aortic valve replacement group (61.3% +/- 6.6% at 8 years) compared with the mitral valve replacement group (54.4% +/- 7.7% at 8 years; p = 0.04). This study confirms the satisfactory performance of the Carpentier-Edwards valve after aortic valve replacement in elderly patients.(ABSTRACT TRUNCATED AT 400 WORDS)
Louagie, Yves ; Noirhomme, Philippe ; Aranguis, E ; Eucher, Philippe ; Vanruyssevelt, P. ; et. al. Use of the Carpentier-Edwards porcine bioprosthesis: assessment of a patient selection policy.. In: The Journal of thoracic and cardiovascular surgery, Vol. 104, no. 4, p. 1013-24 (1992)