After a brief review of the main ideas as the basis of recent commercial automatic impedance bridges, the paper emphasizes the study of the four-projections measurement method. Several potential error sources are considered, as sensitivity limits, quadrature and timing errors. A short presentation is made of a microprocessor assisted laboratory prototype working at any frequency between 10 Hz and 60 kHz by means of an incorporated synthesizer and offering a basis accuracy of 0.2%.
Bibliographic reference |
Eugène, Christian. Automatic impedance bridges. Theoretical and experimental study of the four-projection method.Technological and Methodological Advances in Measurement. Acta IMEKO 1982. Proceedings of the 9th IMEKO Congress of the International Measurement Confederation (Berlin, West Germany, 24-28 May 1982). In: Havrilla, K.; Solt, J.; Kemeny, T.;, Technological and Methodological Advances in Measurement. Acta IMEKO1982. Proceedings of the 9th IMEKO Congress of the InternationalMeasurement Confederation, North-holland1983, p.Vol. 2, p. 331-40 |
Permanent URL |
http://hdl.handle.net/2078.1/68411 |