In order to achieve the harmonization of validation strategies, the interpretation of several validation criteria is proposed. Furthermore, a simple and visual decision tool to assess the validity of an analytical procedure is described: the accuracy profile based on the estimation of the total error of the measurements. This profile build with beta-expectation tolerance intervals can also compute with efficiency the uncertainty related to the results of a laboratory, which is an essential parameter for the accreditation of laboratories under ISO 17025.
Hubert, P. ; Rozet, E. ; Boulanger, Bruno ; Dewe, W. ; Laurentie, M ; et. al. [Synchronization of validation and estimation strategies of doubt associated as part of the accreditation in trial laboratories].3rd Meeting of the Internationales-de-Toxicologie-Hospitaliere (Liege(Belgium), Nov 24-25, 2005). In: Acta Clinica Belgica (Multilingual Edition), Vol. 61, p. 54-56 (2006)