Rouvez, Alice
[UCL]
Herman, Benoît
[UCL]
Kerckhofs, Greet
[UCL]
This thesis presents the design, assembly and test of a bioreactor for a monoaxial muscle, that includes both electrical and mechanical stimulation. A first version of this bioreactor was designed last year (2017) but several remaining issues prevent its proper functioning. Therefore, the transmission and the chamber designs were changed to ensure a biocompatible, airtight, and sterilizable bioreactor, with an easy mounting, and a functional stimulation. The new design showed great results with regard to the sealing and the mounting of the bioreactor, but a new membrane with less rigidity should be found to ensure the actuator shaft sealing. The different implemented stimulation modes were tested first with springs and weight, and afterward on a pork abdominal muscle, showing good results as well. Even if further investigations should be done regarding the filtering of the force measurements, this bioreactor have already brought a lot of improvements.


Référence bibliographique |
Rouvez, Alice. Bioreactor including electrical and mechanical stimulation for a monoaxial muscle. Ecole polytechnique de Louvain, Université catholique de Louvain, 2018. Prom. : Herman, Benoît ; Kerckhofs, Greet. |
Permalien |
http://hdl.handle.net/2078.1/thesis:17244 |